Langmuir Trough

In the summer of 2004, ChemMatCARS successfully commissioned a Langmuir Trough for use in liquid-vapor interface studies. The trough is designed to be used in conjunction with the x-ray liquid spectrometer available in ChemMatCARS station 15-ID-C, but also may be used as a stand-alone instrument in our chemistry lab (schedule permitting). The trough may be used to obtain information about interfacial structure at aqueous solution/vapor interfaces. Langmuir monolayers and thin films at such interfaces are also readily probed. The primary x-ray techniques which may be used with this instrument follow:

  • Reflectivity provides information about film thickness, surface roughness and electron density profile normal to the interface.
  • In-plane diffraction provides information about ordering within the plane of the interface.

Langmuir Trough
Langmuir Trough mounted on the liquid surface spectrometer in station 15-ID-C.
X-rays go from right to left

The Langmuir Trough was custom built for the x-ray reflectivity and grazing incidence measurements described above. Other features include:

  • Temperature control: a Lakeshore, Model 330 is used to control sample temperature, and has the capability for both heating and cooling. (Maximum temperature ~40°C)
  • Sealing: near-hermetic sealing allows for the introduction of helium into the sample environment, thereby minimizing ozone production and x-ray absorption.
  • Software: a widely used package from Nima Technology Inc. controls the barrier motion and Wilhelmy plate pressure sensor readout.
  • Common operational modes: isothermal compressions and constant pressure monitoring.
  • Compression: a compression ratio of 10:1 may be achieved without simultaneous x-ray measurement, and up to 5:1 for measurements with x-rays.
The trough is currently available for general users. It may be used offline under a laminar flow hood in our chemistry laboratory.

Covered Langmuir Trough in laminar flow hood Covered Langmuir Trough in laminar flow hood. Langmuir Trough with top lid removed Langmuir Trough with top lid removed. (Trough is made from teflon. X-rays go through the polyimide windows on the right -barrier has been removed)

Technical Specifications

Dimensions of liquid footprint without the barrier
  • Length: 42.0 cm
  • Width: 8.9 cm
  • Usable Area (no x-rays): 35-335 cm2
  • Usable Area (x-rays): 120-335 cm2 (actual min. area may be reduced slightly more)
  • Depth (no meniscus): 7.0 mm
  • Recommended Fill Volume: 350 mL
  • Barrier speed range: 0.01 – 0.5 mm/s
  • Control software: Nima v5.16

User Help

The following documents are available online for consultation:

Acknowledgments

The following people contributed to the design of this trough: Binhua Lin, David Schultz, C. ‘George’ Pullins (ChemMatCARS), Ka Yee Lee, Maria Ratajczak, GuoHui Wu, Canay Ege, Steve Danauskas, (UC Chemistry) and Mark Schlossman (UIC Physics).


ChemMatCARS • 9700 S. Cass Avenue • Argonne, IL 60439 • Tel. (630) 252-0450 • Fax. (630) 252-0460